82
Check
item
Problem
Details of inspection Corrective action
Reference
page
Computer
output
Computer data is not
output.
Is the baud rate correctly
set?
Is the cable connector
securely connected to the
port?
Has the computer been
turned ON before the
controller?
Is the correct program used?
Set the controller and compu-
ter to the same baud rate.
Note, however, that some
computers may not recognize
correct measurements when
set to a high baud rate.
Check that the cable connec-
tor is securely connected to
the port.
Be sure to turn ON the com-
puter first, then controller.
Send a message such as
"X0" from the computer to
access the controller.
CHAPTER 9 Troubleshooting Guide
p. 27
p. 49
ls3100_76_88.p65 08.10.1, 10:57 AM82